SH-280 auto hematology analyzer

SH-280 auto hematology analyzer
Product Detailed
3 part differentiation of WBC,20 parameters + 3 histograms,auto mode and manual mode are selectable

Specifications

Test principle: electrical resistance for counting and SFT method for HGB

Test flow: fully auto and the blood don’t need dilute or other process

Clot clear: high voltage burn and high pressure flush

Parameter: 3 part differentiation of WBC,20 parameters + 3 histograms

Sample volume: 9.8μl peripheral blood or venous blood

Test channel: two channels

Throughput: 60 samples/hour

Language: English

QC mode: , -R, -B,L-J, auto draw and print QC figure

Calibration mode: auto mode and manual mode are selectable, provide original calibrator and control

Reference range type: newborn,children,male,female and universal

Reagent: provide original reagent

Memory: save more than 30000 sample’s detail results including histograms

Display: Large colour LCD, show all parameters and histograms at same screen

Input and output: keyboard,mouse,Parallel Port,RS232 serial port, can be connected to computer

Printer: Inner thermal printer,can be connected to external computer and printer

Working Environment: Temperature;18°C~35°C;Humidity;40%RH-80%RH

Voltage: AC 175 V~265V,Auto adapt according to the input voltage

Dimension: 300 mm×400 mm×440mm(L×W×H)

Net weight: 18Kg

Performance

Item

Precision, CV

WBC

≤2.0%

RBC

≤1.9%

HGB

≤1.9%

MCV

≤0.4%

PLT 

≤4.0%

Carry over

WBC≤2%,RBC≤1%,HGB≤2%,PLT≤2%

Technical parameter

Sample volume: normal mode: 9.8μl, dilution mode: 20μl

Aperture diameter: WBC: 100μ m, RBC,PLT: 70μ m

Throughput: 60 samples/hour

Clot clear: high voltage burn and high pressure flush

Calibration mode: auto mode and manual mode are selectable, provide original calibrator and control

QC mode: , -R, -B,L-J, auto draw and print QC figure

Display: Large colour LCD, show all parameters and histograms at same screen

Memory: save more than 30000 sample’s detail results including histograms

Input and output: keyboard,mouse,Parallel Port,RS232 serial port, can be connected to computer

Printer: Inner thermal printer,can be connected to external computer and printer

Test parameter

WBC  Lymph#  Mid#  Gran#  Lymph%  Mid%  Gran%  RBC  HGB  HCT

MCV  MCH  MCHC  RDW-SD  RDW-CV  PLT  MPV  PDW  PCT  P-LCR

WBC Histogram  RBC Histogram  PLT Histogram

SH-280 auto hematology analyzer